Test Code 0230214 Venous Blood Gas
Methodology
ABL800 FLEX Blood Gas System by Radiometer
Includes pH, pCO2, pO2, Base Excess, HCO3 (bicarbonate), Total CO2, O2 saturation, HbCO, metHgb and O2 Status.
Performing Laboratory
Alomere Health Hospital Laboratory
Specimen Requirements
Specimen Type: Venous whole blood
Collection Container/Tube: Sodium or lithium heparinized tube
Specimen Volume: 2.0 mL
Specimen Minimum Volume: 2.0 mL
Collection Instructions:
1. The patient should be at rest for at least 20 minutes prior to blood draw.
2. Specimen should be analyzed within 30 minutes of collection.
Specimen Transport Temperature
Ambient
Reference Values
pH
7.32-7.42
Critical values (automatic call-back): ≤7.20 or ≥7.60
pCO2
41.0-51.0 mm Hg
pO2
25.0-40.0 mm Hg
Base Excess
No reference range
HCO3 (Bicarbonate)
24.0-28.0 mmol/L
Critical values (automatic call-back): ≤10.0 or ≥40.0 mmol/L
Total CO2
25.0-29.0 mmol/L
O2 SATURATION
45.0-75.0%
HbCO
≤2.0%
Critical values (automatic call-back): ≥20%
MetHgb
No reference range
Day(s) Test Set Up
Monday through Sunday
Test Classification and CPT Coding
82805