Test Code C5FX C5 Complement, Functional, Serum
Reporting Name
C5 Complement, Functional, SUseful For
Diagnosis of C5 deficiency
Investigation of a patient with an undetectable total complement level
Performing Laboratory
Mayo Clinic Laboratories in RochesterSpecimen Type
Serum RedOrdering Guidance
The total complement assay (COM / Complement, Total, Serum) should be used as a screen for suspected complement deficiencies before ordering individual complement component assays. A deficiency of an individual component of the complement cascade will result in an undetectable total complement level.
Specimen Required
Patient Preparation: Fasting preferred
Supplies: Sarstedt 5 mL Aliquot Tube (T914)
Collection Container/Tube: Red top
Submission Container/Tube: Plastic vial
Specimen Volume: 1 mL
Collection Instructions:
1. Immediately after specimen collection, place the tube on wet ice.
2. Centrifuge and aliquot serum into plastic vial.
3. Immediately freeze specimen.
Specimen Minimum Volume
0.5 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum Red | Frozen | 14 days |
Reference Values
29-53 U/mL
Day(s) Performed
Monday through FridayTest Classification
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.CPT Code Information
86161
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
C5FX | C5 Complement, Functional, S | 60472-8 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
C5FX | C5 Complement, Functional, S | 60472-8 |
Clinical Information
Complement proteins are components of the innate immune system. There are 3 pathways to complement activation: 1) the classical pathway, 2) the alternative (or properdin) pathway, and 3) the lectin (or mannan-binding lectin) pathway. The classical pathway of the complement system is composed of a series of proteins that are activated in response to the presence of immune complexes. A single IgM molecule or 2 IgG molecules are sufficient to trigger activation of the recognition complex initiated by C1q. The activation process triggers a cascade that includes an amplification loop. The amplification loop is mediated by C3, with cleavage of a series of proteins, and results in 3 main end products: 1) anaphylatoxins that promote inflammation (C3a, C5a), 2) opsonization peptides that are chemotactic for neutrophils (C3b) and facilitate phagocytosis, and 3) the membrane attack complex (MAC), which promotes cell lysis.
Patients with deficiencies of the late complement proteins (C5, C6, C7, C8, and C9) are unable to form the MAC and may have increased susceptibility to neisserial infections. Approximately 40 cases of C5 deficiency have been reported.Â
Complement levels can be detected by antigen assays that quantitate the amount of the protein (C5AG / C5 Complement, Antigen, Serum). For most of the complement proteins, a small number of cases have been described in which the protein is present but is nonfunctional. These rare cases require a functional assay to detect the deficiency.
Interpretation
Low levels of complement may be due to inherited deficiencies, acquired deficiencies, or due to complement consumption (eg, as a consequence of infectious or autoimmune processes).
Absent C5 levels in the presence of normal C3 and C4 values are consistent with a C5 deficiency. Absent C5 levels in the presence of low C3 and C4 values suggest complement consumption.
Normal results indicate both normal C5 protein levels and normal functional activity.
Report Available
1 to 3 daysSpecimen Retention Time
14 daysReject Due To
Gross hemolysis | OK |
Gross lipemia | Reject |
Gross icterus | OK |
Method Name
Automated Liposome Lysis Assay