Sign in →

Test Code FLEC Flecainide, Serum

Reporting Name

Flecainide, S

Useful For

Optimizing flecainide dosage

 

Assessing flecainide toxicity

 

Monitoring compliance

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Serum Red


Specimen Required


Supplies: Sarstedt Aliquot Tube, 5 mL (T914)

Collection Container/Tube: Red top (serum gel/SST are not acceptable)

Submission Container/Tube: Plastic vial

Specimen Volume: 1.5 mL

Collection Instructions:

1. Draw blood immediately before next scheduled dose.

2. Centrifuge and aliquot serum into a plastic vial within 2 hours of collection.


Specimen Minimum Volume

0.5 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Serum Red Refrigerated (preferred) 28 days
  Ambient  28 days
  Frozen  28 days

Reference Values

Trough Value

0.2-1.0 mcg/mL: Therapeutic concentration

>1.0 mcg/mL: Toxic concentration

Day(s) Performed

Monday through Friday

Test Classification

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information

80181

LOINC Code Information

Test ID Test Order Name Order LOINC Value
FLEC Flecainide, S 3638-4

 

Result ID Test Result Name Result LOINC Value
9243 Flecainide, S 3638-4

Clinical Information

Flecainide (Tambocor) is a Class I cardiac antiarrhythmic agent indicated for treatment of paroxysmal supraventricular dysrhythmia, paroxysmal atrial fibrillation/flutter, and life-threatening ventricular dysrhythmias. After oral administration, flecainide is almost completely absorbed and peak concentrations are attained in approximately 3 hours. The half-life averages approximately 20 hours but is widely variable (12 to 27 hours), and steady-state concentrations are typically achieved in approximately 5 days. Flecainide is eliminated from blood by hepatic metabolism, as well as renal clearance; significant changes in either organ system will cause impaired clearance. Common adverse effects include dizziness, visual disturbances, and dyspnea. Mild-to-moderate toxicity is associated with dizziness, visual disturbances, headache, nausea, fatigue, palpitations, and chest pain. Visual hallucinations and dysarthria may occur at toxic serum concentrations. Death can occur from hypotension, respiratory failure, and asystole.

Interpretation

Flecainide is most effective in premature ventricular contractions suppression at serum concentrations in the range of 0.2 to 1.0 mcg/mL.

 

Serum concentrations above 1.0 mcg/mL are associated with a high rate of cardiac adverse experiences such as conduction defects or bradycardia.

Report Available

2 to 5 days

Specimen Retention Time

14 days

Reject Due To

Gross hemolysis OK
Gross lipemia OK
Gross icterus OK

Method Name

Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)

Forms

If not ordering electronically, complete, print, and send a Therapeutics Test Request (T831) with the specimen.